| General |
| Read method |
Endpoint, kinetic, spectral scanning, well area scanning |
| Microplate types |
6-, 12-, 24-, 48-, 96- and 384-well plates, Take3 plate |
| Absorbance |
| Wavelength range |
200 nm - 999 nm, selectable in 1 nm increments |
| Bandpass |
5 nm |
| Dynamic range |
0 to 4.0 OD |
| Resolution |
0.0001 |
| Monochromator wavelength accuracy |
±2 nm |
| Monochromator wavelength repeatability |
±0.2 nm |
| OD accuracy |
0.0 to 2.0 OD: ±1% ±0.010 OD 2.0 to 2.5 OD: ±3% ±0.010 OD |
| OD linearity |
0.0 to 2.0 OD: ±1% ±0.010 OD 2.0 to 2.5 OD: ±3% ±0.010 OD |
| OD repeatability |
0.0 to 2.0 OD: ±1% ±0.005 OD 2.0 to 2.5 OD: ±3% ±0.005 OD |
| Reading speed |
96-well Normal: 49 seconds 96-well Rapid: 38 seconds 96-well Sweep: 15 seconds |
| Physical Characteristics |
| Dimensions |
12”W x 12.5”D x 7.7”H (30.5 x 31.7 x 19.5 cm) |
| Weight |
14 lbs (6.35 kg) |
| Regulatory |
| Regulatory |
For in vitro diagnostic use. All BioTek microplate instrumentation is CE and ETL marked. For details on approvals and standards compliance, please contact BioTek.
Compliance: EN 61326-1:2006;EN61326-2-6:2006;EN61000-4-2; EN61000-4-3, EN61000-4-4; EN61000-4-5; EN61000-4-6; EN61000-4-11 |