Performance characteristics
| Synonyms |
Glu |
| Sample type |
Serum,plasma |
| Sensitivity |
0.05 mmol/L |
| Detection range |
0.05-30 mmol/L |
| Detection method |
Colorimetric method |
| Assay type |
Quantitative |
| Assay time |
60 min |
| Precision |
Average inter-assay CV: 1.300%Average intra-assay CV: 1.200% |
| Other instruments required |
Vortex mixer, Micropipettor, Incubator |
| Other reagents required |
Normal saline (0.9% NaCl) |
| Storage |
2-8℃ |
| Valid period |
12 months |
Dilution of sample
It is recommended to take 2~3 samples with expected large difference to do pre-experiment before formal experiment and dilute the sample according to the result of the pre-experiment and the detection range (0.05-30 mmol/L).
The recommended dilution factor for different samples is as follows (for reference only):
|
Sample type
|
Dilution factor
|
|
Rat plasma
|
1
|
|
Mouse serum
|
1
|
|
Human serum
|
1
|
|
Human plasma
|
1
|
Note: The diluent is normal saline (0.9% NaCl).