Набор для колориметрического анализа железа / Iron Colorimetric Assay Kit
- Detection instrument: Spectrophotometer(520 nm)
- Detection method: Colorimetric method
Manual
Detection principle
Under the action of acidic solution and reductant, ferric ions can be separated from transferrin in serum, and reduced into ferrous ions (Fe2+). The latter then bind to bipyridine and form pink complexes. The concentration of iron can be calculated by measuring the OD value at 520 nm indirectly.
Performance characteristics
| Synonyms |
Fe |
| Sample type |
Serum,tissue |
| Sensitivity |
0.072 mg/L |
| Detection range |
0.072-60 mg/L |
| Detection method |
Colorimetric method |
| Assay type |
Quantitative |
| Assay time |
50 min |
| Precision |
Average inter-assay CV: 2.500%Average intra-assay CV: 4.600% |
| Other instruments required |
Micropipettor, Vortex mixer, Incubator, Centrifuge |
| Other reagents required |
Normal saline (0.9% NaCl), PBS (0.01 M, pH 7.4) |
| Storage |
2-8℃ |
| Valid period |
12 months |